Center for Biofilm Engineering
Abstract:
"Improving the AOAC use-dilution method by establishing a minimum log
density value for test microbes on inoculated carriers"
09-027
The AOAC Use-Dilution methods, 955.14 (Salmonella enterica), 955.15 (Staphylococcus
aureus), and 964.02 (Pseudomonas aeruginosa), are used to measure the
efficacy of disinfectants on hard inanimate surfaces. The methods do not provide
procedures to assess log density of the test microbe on inoculated penicylinders
(carrier counts). Without a method to measure and monitor carrier counts, the
associated efficacy data may not be reliable and repeatable. This report
provides a standardized procedure to address this method deficiency. Based on
carrier count data collected by four laboratories over an 8 year period, a
minimum log density value is proposed to qualify the test results. Carrier count
data were collected concurrently with 242 Use-Dilution tests. The tests were
conducted on products bearing claims against P. aeruginosa and S.
aureus with and without an organic soil load (OSL) added to the inoculum (as
specified on the product label claim). Six carriers were assayed per test for a
total of 1452 carriers. All 242 mean log densities were at least 6.0 (geometric
mean of 1.0´ 106 CFU/carrier). The mean log densities did not exceed 7.5
(geometric mean of 3.2 ´ 107 CFU/carrier). For all microbes and OSL treatments,
the mean log density (± SEM) was 6.7 (± 0.07) per carrier (a geometric mean of
5.39 ´ 106 CFU/carrier). The mean log density for six carriers per test showed
good repeatability (0.29) and reproducibility (0.32). A minimum mean log density
of 6.0 is proposed as a validity requirement for S. aureus and P.
aeruginosa. The minimum level provides for the potential inherent
variability that may be experienced by a wide range of laboratories and the
slight effect due to the addition of an OSL. A follow-up report is planned to
present data to support the carrier count procedure and carrier counts for S.
enterica.
Tomasino SF, Pines RM, Hamilton MA, "Improving the AOAC use-dilution method by
establishing a minimum log density value for test microbes on inoculated
carriers," J AOAC International 2009; 92(5):1531-1540
|